[Poster Board #E3] Non-Invasive Measurements of Carboxyhemoglobin and Methemoglobin in Pediatric Patients with Sickle Cell Disease, [Publication Page: A4788]
Bland and Altman analysis was used to assess agreement between measurement techniques. The Radical-7 bias was -0.34% for COHb and -0.03% for MetHb. The precision, an indicator of measurement uncertainty, was 1.30% for COHb and 0.48% for MetHb. Differences in measurements between the two techniques were all within the 95% confidence interval of the bias, -2.9% to 2.2% for COHb and -0.96% to 0.90% for MetHb, and within the published uncertainty of the Radical-7 (COHb: 3% and MetHb: 1%).Conclusions. These results demonstrate that a non-invasive co-oximeter is useful in measuring COHb and MetHb levels in pediatric patients with SCD. Although the non-invasive slightly underestimated the invasive technique, there was close agreement between the two. Am J Respir Crit Care Med 179;2009:A4788 Session Info: [**] Thematic Poster Session, [C60] CLINICAL AND EPIDEMIOLOGICAL STUDIES Day/Date: Tuesday, May 19, 2009 Session Time: 8:15 AM - 4:00 PM Poster Viewing: 10:45-12:30 Room: Area E (Sails Pavilion, Upper Level), SAN DIEGO CONVENTION CENTER
|